Vertical Sliding Doors Hospital Steam Sterilizer With SS304 Full Jacket Chamber
Key features:
Application
General-purpose steam sterilization of surgical instruments, textiles and hospital utensils in central sterilization departments, operation departments and laundries. The temperature range is from
104°C–135°C.
Technical chart:
Model | Chamber Size | Overall Size | Net Weight | Steam | Water | Power |
W*H*D mm | W*H*D mm | Kg | Kg/cycle | Kg/cycle | KVA | |
MD-0.11V | 400*400*700 | 1000*1800*1040 | 650 | 14 | 215 | 2 |
MD-0.16V | 400*400*1000 | 1000*1800*1340 | 700 | 16 | 220 | 2 |
MD-0.20V | 400*400*1300 | 1000*1800*1640 | 750 | 18 | 225 | 2 |
MD-0.24V | 600*600*670 | 1200*2000*1010 | 800 | 20 | 230 | 2 |
MD-0.30V | 600*600*840 | 1200*2000*1180 | 850 | 22 | 235 | 2 |
MD-0.36V | 600*600*1000 | 1200*2000*1340 | 900 | 25 | 240 | 2 |
MD-0.45V | 600*600*1250 | 1200*2000*1590 | 980 | 27 | 250 | 2 |
MD-0.60V | 600*600*1670 | 1200*2000*2010 | 1100 | 30 | 260 | 3 |
Programs:
Test programs: Bowie & Dick test, vacuum test.
Working programs: Multiple customizable programs for textiles, instruments, rubbers, culture medium and liquids in unsealed bottles.
Main adjustable parameters:
Sterilization temperature rage: at 105°C to 134°C
Sterilization time rage: 0---99 minutes
Vacuum times: 0---9 times
Drying time: 0---99 minutes
Multiple adjustable parameters to ensure user’s customized programs are available, accurate and
perfect sterilization effect.
Sterilization Cycles
Cycle No. | Cycle Name | Further Details |
1 | Instruments/Tools | 3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min. |
2 | Clothing | 3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min. |
3 | Liquid | 15 min exchange, expose temperature 121℃, expose time 15 min, cooling time 25 min. |
4 | B-D Test | 3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min. |
5 | Vacuum Leakage Test | Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect. |
6 | User Programmable | All preset parameters can be resettable; Multiple customized programs can be set and stored by user. |
Watch Video